6. CLEANING AND DISINFECTION

Both the camera head and camera console may be cleaned, but not sterilized. Remove from power
source before any cleaning is performed.
CAUTION: Never immerse or sterilize the camera console as this will damage the camera
console and void the warranty.
The camera head, camera console and headband can be wiped down with commercially available
cleansers commonly used for disinfection of electronic equipment in hospitals such as ethyl or isopropyl
alcohols, disinfecting sprays containing quaternary ammonium compounds, or hydrogen peroxide. The
lenses should be cleaned only with lens tissue, available in any camera store; follow the directions on
the package.
Do not use strongly caustic or acidic cleansers such as "Clorox" hypochlorite bleach,
ammonia, muriatic acid or similar products. Do not use acetone, methyl ethyl ketone or
halogenated / chlorinated hydrocarbon solvents or cleansers containing any of these
restricted compounds.
Apply cleaning agents by light spray or dampened towels.
Do not pour liquids onto the device. Do not allow frank liquids to enter either the camera
head or camera console seams or ventilation openings.
Follow all applicable blood borne pathogen procedures as required by OSHA and/or your hospital, when
cleaning and disinfecting the product.