5.9 FAULT DISPLAYS
There are two fault conditions detectable by sensors that will render the Camera System inoperable
until they are corrected. A fault condition notice will appear on the screen identifying the fault and
how to correct it.
FAN FAULT
In the event of a fan problem the unit will display this screen.
Turn off and restart the Camera System. If the fault persists the unit needs servicing.
SYSTEM IS COOLING
If the internal air temperature exceeds preset limits this screen will display. The fan will run to
reduce the temperature. If acceptable reduction is achieved within ten seconds, the system will turn
back on to resume normal operation.
LIT-239 SUNOPTIC SURGICAL
Rev. B
®
Page 18 of 210
(English)