Setting up the SIRIUS Multigas Detector
5.3 Instrument Alarm Bypass Options
The SIRIUS Multigas Detector (with software version 1.1 or higher) is equipped with a
feature to disable or silence the visual, backlight, and audible options. If any of these
options are disabled during instrument start up, the SIRIUS Detector displays:
•
"VISUAL OFF" if the red LEDs are disabled
•
"AUDIBLE OFF" if the audible buzzer is disabled
•
"BACKLITE OFF" if the backlight is disabled
•
"BACKLITE TIME"
If the visual or audible options are disabled, "ALARM OFF" flashes on the LCD during
Normal Measure mode.
1. Instrument Options Setup
•
Safe LED ON/OFF
•
Operating beep ON/OFF
•
STEL/TWA ON/OFF
2. CAL Lockout Enable
•
To disable calibration in the user mode, set CAL LOCK MODE to ON
•
When ON, calibration is only accessible via the Setup mode and password
(if enabled)
3. CAL DUE Alert
•
To disable CAL Due messages, turn this feature OFF.
•
When ON, the number of days (1 to 180) between calibrations can be set.
User will be required to acknowledge an overdue calibration at turn ON.
4. Warm Up Info
•
Turning this selection OFF causes the instrument NOT to display alarm set
points at power-ON
•
Time (if data logging option installed)
•
Date (if data logging option installed)
5. LEL/CH
Setup
4
•
Sensor ON/OFF (turns the sensor ON or OFF)
•
Display Combustible Gas Type?
•
Methane
•
Pentane
•
Hydrogen
•
Propane
•
LEL or CH
•
Low Alarm (sets the low combustible alarm)
•
High Alarm (sets the high combustible alarm)
•
Cal Gas (sets the expected combustible calibration gas)
30
mode displays % LEL (for any gas) or % CH
4
(for Methane only)
4