What Do I Do If - Seca 757 Mode D'emploi Et Garantie

Masquer les pouces Voir aussi pour 757:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 83

9. WHAT DO I DO IF ...?

Fault
... no weight is dis-
played when scale is
loaded?
... 0.000 does not appear
before weighing?
... a segment lights up
continuously or not at
all?
...
is displayed?
... the display "bAtt"
appears?
... the display "StOP"
appears?
... the display "tEMP"
appears?
... after switching on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
Cause/Remedy
The scale is not connected to the power supply.
- Check if the scale is switched on.
- Check whether the battery block is inserted
and that the plug contacts are properly con-
nected.
A load was placed on the scale before it was
switched on.
- Remove load.
- Switch scale off and back on again.
The relevant segment is faulty.
- Please notify maintenance service.
Battery block power is running down.
- Charge battery block.
Battery block discharged.
- Charge battery block.
The maximum load has been exceeded.
- Remove load.
The ambient temperature of the scale is too high
or too low.
- Set up scale in an ambient temperature
between +10 °C and +40 °C.
- Wait approx. 15 minutes until scale has
adjusted to ambient temperature.
The device was unable to send measured results
to the wireless receiver (seca wireless printer or
PC with seca USB wireless adapter).
– Check that the scale is integrated in the
wireless network.
– Check that the receiver is switched on.
Nearby HF equipment (e.g. mobile phones) are
interfering with reception.
- Make sure that HF equipment is kept at least
1 metre away from transmitters and receivers
in the seca wireless network.
NOTE
If such interference is not eliminated, no fur-
ther acoustic warnings are given with sub-
sequent attempts at transmission.
English
What do I do if ...? • 75

Publicité

Table des Matières
loading

Table des Matières