8. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged.
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.

9. Technical data

Maximum load
Minimum load
Graduations
Dimensions
height:
width:
depth:
Seat height:
Height of figures
Weight
Temperature range
Power supply
Batteries
Power consumption
Calibration as per Directive
2009/23/EC:
Medical product as per Directive
93/42/EEC
EN 60 601-1:
Insulated equipment, Protection Class II:
Accuracy at first calibration
Functions
Model 956
200 kg
2 kg
100 g
906 mm +35,7 mm
563 mm
946 mm
520 mm
20 mm
approx. 22.4 kg
+10 °C to +40 °C
batteries
6 type AA 1.5 V batteries
approx. 25 mA
approved for medical use, Class III
Class I
Range 2-50 kg: ±50 g
Range 50-200 kg: ±100 g
TARE, HOLD
21