6003 • Kronner Manipujector
Instructions for Use • English (continued)
PRECAUTIONS
Test inflate intrauterine balloon prior to insertion.
Lubricate distal end of tube and intrauterine balloon before insertion.
The cervical os should be #13-14 Hank size before inserting the Kronner
Manipujector
tearing the intrauterine balloon.
When injecting any liquid media, closely follow the manufacturer's Directions for
Use that accompanies that product.
POTENTIAL ADVERSE REACTIONS
The following adverse reaction have been suspected or reported. The order of listing
does not indicate frequency or severity.
Injury to uterus (perforation)
Cramping
Infection
Uterine spasm with temporary physiologic blockage of patient fallopian tubes
INSTRUCTIONS FOR USE
1.
Remove the sterile Kronner Manipulator (Uterine Manipulator-Injector) from its
protective peel-apart package. Draw 10 cc of air into a standard plastic syringe and
then insert syringe into the inflation valve assembly. Test inflate the intrauterine
balloon by injection of the air in the syringe. Remove the syringe and check that the
balloon remains inflated.
2.
Following test inflation, reinsert the syringe firmly into the inflation valve assembly
to open the valve. Completely evacuate all the air in the balloon with the syringe.
Then remove the syringe.
3.
With the patient in the lithotomy position, use a vaginal speculum to expose the
uterine cervix and grasp the anterior cervical lip with a single tooth tenaculum.
4.
Probe the uterus for depth and direction with a uterine sound. DO NOT use the
Kronner Manipujector (Uterine Manipulator-Injector) as a uterine sound. Before
inserting the Kronner Manipujector (Uterine Manipulator-Injector) into the uterus,
determine its direction and set the device for effective uterine depth as indicated
by sounding.
To Set the Depth—
A
Locate the depth marks (5 cm thru 10 cm) at the rear underside of the
catheter tube.
B
Firmly squeeze the pinch clamp at the proximal end of the handle to reposition
the catheter within the handle.
C
With the pinch clamp compressed, slide the catheter until the selected depth
mark aligns with the rear edge (I) of the handle marked by the arrow; e.g., if
the uterine depth measures 7 cm, then set the handle at the 7 cm mark.
(Uterine Manipulator-Injector) for easy passage and to prevent
®
Uterine Manipulator-Injector
®
3