9. TROUBLESHOOTING
Problem
No weight displayed
with a load on the scale.
0.0 does not appear
before weighing.
One segment is illumi-
nated either continu-
ously or not at all.
The display
appears.
The display "bAtt"
appears.
The display "StOP"
appears.
The display "tEMP"
appears.
►
Have servicing and repairs carried out exclu-
sively by seca service or by an authorized ser-
vice partner.
►
Please see the section entitled "Servicing/re-
calibration" on page 107.
Cause / Solution
The scale has no power supply.
- Check whether the scale is switched on.
A load was already on the scale before it was
switched on.
- Remove the load from the scale.
- Switch the scale off and back on again.
The corresponding point has a fault.
- Inform the service department.
Battery block power is running down.
- Recharge battery block as soon as possible.
Battery block discharged.
- Charge battery block.
Maximum load has been exceeded.
- Remove the load from the scale.
The ambient temperature of the scale is too high
or too low.
- Set up the scale in an ambient temperature
between +10 °C and +40 °C.
- Wait for around 15 minutes until the device
has adapted to the ambient temperature.
English
Troubleshooting • 105