10.TROUBLESHOOTING
Problem
No weight displayed
with a load on the scale.
0.0 does not appear
before weighing.
One segment is illumi-
nated either continu-
ously or not at all.
The display
appears.
The display "bAtt"
appears.
The display "StOP"
appears.
The display "tEMP"
appears.
94 •
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from page 94.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 94, you may not use the device.
– Have the device repaired by seca service or
by an authorized service partner.
– Please see the section entitled
"Servicing/recalibration" on page 96.
Cause / Solution
The scale has no power supply.
- Check whether the scale is switched on.
- Check whether batteries are inserted.
A load was already on the scale before it was
switched on.
- Remove the load from the scale.
- Switch the scale off and back on again.
The corresponding point has a fault.
- Inform the service department.
Battery power is running down.
- Insert new batteries as soon as possible.
Batteries are flat.
- Insert new batteries.
Maximum load has been exceeded.
- Remove the load from the scale.
The ambient temperature of the scale is too high
or too low.
- Set up the scale in an ambient temperature
between +10 °C and +40 °C.
- Wait for around 15 minutes until the device
has adapted to the ambient temperature.