order code
measurement range
linearity
(relative)
1)
cell constant
2)
max. fluid temperature
fluid pressure
temperature sensor
electrical connection
± 0.5% accuracy can be achieved when calibration is performed in
1)
a conductivity range close to that of the process. With an extended
measuring range, the linearity is about 5%. The linearity also
depends on the device used.
Value measured according to the Bürkert standard procedure. The
2)
cell constant may be influenced by the mounting position.
8
8221 with 4 electrodes and 1,5"
clamp connection
short insertion depth
557719
0.1 µS/cm... 500 mS.cm
-1
±0.5... 5%
0.147 cm
-1
-20 to 135 °C. max
6 bar max.
Pt1000
High temperature cable, length 5 m,
instrument side with open wire.
English
Type 8221
Technicaldata
order code
measurement range
linearity
(relative)
1)
cell constant
2)
max. fluid temperature
fluid pressure
temperature sensor
electrical connection
± 0.5% accuracy can be achieved when calibration is performed in
1)
a conductivity range close to that of the process. With an extended
measuring range, the linearity is about 5%. The linearity also
depends on the device used.
Value measured according to the Bürkert standard procedure. The
2)
cell constant may be influenced by the mounting position.
8221 with 4 electrodes and 1,5"
clamp connection
long insertion depth
558884
0.1 µS/cm... 500 mS.cm
-1
±0.5... 5%
0.147 cm
-1
-20 to 135 °C. max
6 bar max.
Pt1000
High temperature cable, length 5 m,
instrument side with open wire.