4.2 Remote control commands
90
Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the end of the
determination can be switched off. Thus, for example, in conjunction
with a Coulometer several determinations can be performed in the same
beaker. Before AutoEOD is switched on, line 3 must be set to "OFF".
.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inactive"
means resetting of the signal, "pulse" means output of a pulse of app.
150 ms, "OFF" means the line is not operated, see also page 131.
Warnings:
• If you have "AutoEOD" to "ON", an active line 3 is set to "inactive" by
the EOD pulse.
• L6 is the line of the activate pulse. An active line 6 is set to "inactive"
by the activate pulse.
• L5 is the error line. It is continuously controlled by the Coulometer pro-
gram and can therefore not be set freely.
Line assignments in Coulometer program:
L0
Ready, inactive state
L1
Conditioning OK
L2
Titration in progress
L3
EOD (End Of Determination)
L4
---
L5
Error
L6
Activate pulse + can be set in TIP
L7
Pulses for recorder
L8
Connected remote box
L9,10
---
L11
Change reagent
L12
Sample size out of limits
L13
Result out of limits
.ResetLines: Lines are set to the inactive status (= high).
4.2.2.80. Assembly.Stirrer.Status
Switching stirrer ON/OFF.
4.2.2.81. Assembly.Bur.Empty
Assembly.Prep
Starts the function "empty" and "preparation" resp. on the connected Dosino.
4.2.2.82. Assembly.Bur.Rates.Forward.Selected
Assembly.Bur.Rates.Forward.Digital
Assembly.Bur.Rates.Reverse.Selected
Assembly.Bur.Rates.Reverse.Digital
Expelling and aspirating rate in mL/min. "max." means maximum possible rate with
the Exchange Unit in current use.
4.2.2.83. Assembly.Bur.Fill
ON, OFF
$G, $S, $H, $C
$G, $S, $H, $C
0...150, max.
0...150, max.
$G, $H, $C
756/831 KF Coulometer, Instructions for Use