.UnitFlow: Unit of gas flow.
4.2.2.74. Info.ActualInfo.Display.L1
Lines of the display. The display can be written to from the computer. Proceed as
follows:
1. Lock the display, see 4.2.2.90.
2. Delete the whole display (.DelAll).
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 4.2.2.90
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 4.2.2.47) to refresh the dis-
play.
$Q sends the contents of the corresponding display line.
4.2.2.75. Info.ActualInfo.Comport.Number
$Q sends the comport number of the Coulometer where the PC is connected.
4.2.2.76. Info.Assembly.CycleTime
Inquiries regarding basic variables of the assembly.
.Cycle time: Time of measuring cycles in s (0.4).
.ExV:
.DeviceTemp: Internal temperature of Coulometer in °C.
4.2.2.77. Assembly.GenEl.Pulse
Control of the generator electrode. The pulse will be generated with &A.G.P$G.
.Length:
.Current:
4.2.2.78. Assembly.Meas.Status
Control of the indicator electrode. When the measuring function is switched on, no
method can be started at the Coulometer.
.Ipol:
4.2.2.79. Assembly.Outputs.AutoEOD
756/831 KF Coulometer, Instructions of Use
Info.ActualInfo.Display.L8
Info.ActualInfo.Display.DelAll
Info.Assembly.ExV
Info.Assembly.DeviceTemp
Volume of the Dosing Unit of the connected Dosino in mL.
Assembly.GenEl.Pulse.Length
Assembly.GenEl.Pulse.Current
Length of pulse in 2000 steps. 2000 means a pulse of 400 ms (e.g. a
pulse of 150 ms would mean 750 steps).
Current for pulse in mA.
Assembly.Meas.Ipol
Polarization current in uA.
Assembly.Outputs.SetLines
Assembly.Outputs.SetLines.L0
up to .L13
Assembly.Outputs.ResetLines
4.2 Remote control commands
up to 32 ASCII characters
up to 32 ASCII characters
read only
read only
read only
read only
0...2000
0, 100, 200, 400
ON, OFF
2, 10, 20, 40
ON, OFF
active, inactive, pulse, OFF
$G
$G
$G
$G
89