4.2 Remote control commands
86
C40: Initial measured value in mV, e.g. "226".
C41: Mass of water in ug, e.g. "126.5"
C42: Time from start of titration to end in s, e.g. "26".
C43: Drift at titration start in ug/min, e.g. "5.1".
C44: Titration temperature in °C, e.g. "25.0"
C45: Total charge in mA·s, e.g. "1355.5"
4.2.2.67. Info.StatisticsVal.ActN
Info.Statistics.1.Mean
Info.Statistics.1.Std
Info.Statistics.1.RelStd
etc. up to .9
The current values of the statistics calculation.
$Q sends, e.g.
ActN: Current value of the individual results
Data for MN1:
Mean: Mean value (decimal places as in result)
Std: Standard deviation (1 decimal place more than in result)
RelStd: Relative standard deviation (in %, 2 decimal places)
4.2.2.68. Info.SiloCalc.C24.Name
Info.SiloCalc.C24.Value
Info.SiloCalc.C24.Unit
for .C25 as for .C24
Info.SiloCalc.C26.ActN
Info.SiloCalc.C26.Mean
Info.SiloCalc.C26.Std
Info.SiloCalc.C26.RelStd
for .C27 as for .C26
The current values from the silo calculations. C26 is the mean value out of the C24
variables; C27 comes from C25.
$Q sends:
C24.Name: Name of the assigned value
C24.Value: Value
C24.Unit: Unit of the assigned value
C26.ActN: Number of single results
C26.Mean: Mean (decimal places as for the result itself)
C26.Std: Standard deviation (decimal places as for the result + 1)
C26.RelStd: Relative standard deviation (in %, 2 decimal places)
4.2.2.69. Info.ActualInfo.Inputs.Status
Info.ActualInfo.Inputs.Change
Info.ActualInfo.Inputs.Clear
Info.ActualInfo.Outputs.Status
Info.ActualInfo.Outputs.Change
Info.ActualInfo.Outputs.Clear
Status sends the current status of the I/O lines, Change sends the information re-
garding whether a change in status of a line has taken place since the last clearing,
read only
read only
read only
read only
"0.0231"
read only
read only
read only
read only
read only
read only
read only
"0.0231"
read only
read only
read only
read only
756/831 KF Coulometer, Instructions for Use
"3"
"3.421"
"0.14"
"RS1"
"2.222"
"%"
"3"
"3.421"
"0.14"
$G
$G