4.2 Remote control commands
78
Mode.Def.Formulas.1.Decimal
Mode.Def.Formulas.1.Unit
Mode.Def.Formulas.1.Limits
Mode.Def.Formulas.1.LoLim
Mode.Def.Formulas.1.UpLim
Mode.Def.Formulas.1.Output
Mode.Def.Formulas.2.Formula
etc. up to .9
Entry of formulas. Rules for formula entry, see page 34.
Example: "H2O*C01/C00"
In addition to the formula, a text for result output, the number of decimal places and
a unit for the result output can be selected. "No unit" is selected with the blank string.
In place of "RSX", a result name may be entered (.TextRS). This name is outputted in
the result report, scalc full and scalc srt. It is used for the result and the correspond-
ing mean value.
The limit control for results can also be activated. If a result is out of limit, a message
appears in the result report, E196 is sent, and output line L13 can be set.
4.2.2.23. Mode.Def.SiloCalc.Assign.C24
Mode.Def.SiloCalc.Assign.C25
Mode.Def.SiloCalc.MatchId
.Assign.C2X: Assignment to store results in the silo as C2X.
.MatchId:
Indication which sample identification(s) have to match so that the
results can be combined.
4.2.2.24. Mode.Def.ComVar.C30
Mode.Def.ComVar.C31
etc., up to .C39
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They can
be viewed and entered there, see 4.2.2.54.
4.2.2.25. Mode.Def.Report.Internal (only 756)
Mode.Def.Report.Assign1
Mode.Def.Report.Assign2
Definition of the report sequence, which is outputted automatically at the end of the
determination. Entries of more than one block have to be separated with ";".
.Internal:
Internal printer of the Coulometer. (only 756)
.Assign1: Output to COM1 of the Coulometer.
.Assign2: Output to COM2 of the Coulometer.
4.2.2.26. Mode.Def.Mean.1.Assign
Mode.Def.Mean.2.Assign
etc., up to .9
Assignment of the statistics calculations. Valid assignments are a requirement for
statistics calculations. In addition, the statistics calculation must be switched on, see
4.2.2.13. Rules for statistics calculations see page 37.
meas crv, comb, mplist, param, calc, scalc full, scalc srt, ff
756/831 KF Coulometer, Instructions for Use
0...2...5
6 ASCII characters
ON, OFF
0...±999 999
0...±999 999
active, pulse, OFF
RSX, H2O, CXX
RSX, H2O, CXX
id1, id1&2, all, OFF
RSX, MNX, H2O, CXX
result, water crv, rate crv,
ditto
ditto
RS1, RSX, H2O, CXX

Publicité

Table des Matières