Selection of the values from Assembly for the output in the set time interval (see
4.2.2.92):
.CyclNo:
.I:
.Meas:
.Pot:
.IPulse:
.Bur.V:
The unit "assembly" must be preset (see 4.2.2.92).
4.2.2.94. Setup.SendMeas.Titrator.CyclNo
Selection of the values from the titrator which are sent in the set time interval (see
4.2.2.91):
.CyclNo:
.Water:
.Meas:
.dWaterdt: Rate or drift associated to the cycle number in ug/min, e.g. "23.0".
.I:
.Pot:
.IPulse:
OV will be sent for "overrange".
The unit "titrator" must be preset (see 4.2.2.92).
4.2.2.95. Setup.AutoInfo.Status
756/831 KF Coulometer, Instructions of Use
Setup.SendMeas.Assembly.Pot
Setup.SendMeas.Assembly.IPulse
Setup.SendMeas.Assembly.Bur.V
Cycle number of the potential measurement. Together with the cycle
time (4.2.2.77), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and it is al-
ways incremented as long as the instrument remains switched on.
Total charge in mA·s associated to the cycle number, e.g. "667.48".
Measured value in mV associated to the cycle number, e.g. "104.2".
Voltage at generator electrode associated to the cycle number.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means
>28 V.
Current of pulse associated to the cycle number.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
Dosed volume of connected Dosino in ml, e.g. "5.234".
Setup.SendMeas.Titrator.Water
Setup.SendMeas.Titrator.Meas
Setup.SendMeas.Titrator.dWaterdt
Setup.SendMeas.Titrator.I
Setup.SendMeas.Titrator.Pot
Setup.SendMeas.Titrator.IPulse
Cycle number. Together with the cycle time (4.2.2.78), a time frame can
be set up. The other data belong to the corresponding cycle number. The
cycle number is set to 0 at the start of a method and it is incremented
until the end of the method.
Total water associated to the cycle number in ug, e.g. "62.313"
Measured value in mV at the indicator electrode associated to the cycle
number, e.g. "104.2".
Total charge in mA·s associated to the cycle number, e.g. "667.48".
Voltage at generator electrode associated to the cycle number.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means
>28 V.
Current of actual pulse associated to the cycle number.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
Setup.AutoInfo.P
Setup.AutoInfo.T.R
Setup.AutoInfo.T.G
Setup.AutoInfo.T.GC
4.2 Remote control commands
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
93