4.2 Remote control commands
88
Info.ActualInfo.Titrator.I
Info.ActualInfo.Titrator.Pot
Info.ActualInfo.Titrator.IPulse
$Q sends the current values.
.CyclNo:
Cycle number of the voltage measurement cycle, e.g. "127". From the
cycle number and the cycle time (see 4.2.2.77), a time frame can be set
up. The cycle number is set to 0 on switching on the instrument and on
every start. It is incremented as long as the instrument remains switched
on.
.Water:
Total water in ug, e.g. "62.313"
.Meas:
Measured value in mV at the indicator electrode, e.g. "104.2".
.dWaterdt: Rate or drift in ug/min, e.g. "23.0".
.I:
Total charge in mA·s, e.g. "667.48".
.Pot:
Voltage at generator electrode.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means
>28 V.
.IPulse:
Current of actual pulse.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
OV will be sent for "overrange".
4.2.2.72. Info.ActualInfo.MeasPt.Index
Info.ActualInfo.MeasPt.X
Info.ActualInfo.MeasPt.Y
Info.ActualInfo.MeasPt.Z1
Info.ActualInfo.MeasPt.Z2
Info.ActualInfo.EP.Index
Info.ActualInfo.EP.X
Info.ActualInfo.EP.Y
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry into
the list of EP.
.MeasPt.X Time in s, e.g. "14".
.MeasPt.Y Water in ug, e.g. "27.5".
.MeasPt.Z1 Measured value in mV, e.g. "160.3".
.MeasPt.Z2 Rate in ug/min, e.g. "100.5".
.EP.X
Water in ug, e.g. "26.6".
.EP.Y
Measured value in mV, e.g. "98.6".
4.2.2.73. Info.ActualInfo.Oven.HeatTime
Info.ActualInfo.Oven.SampleTemp
Info.ActualInfo.Oven.LowTemp
Info.ActualInfo.Oven.HighTemp
Info.ActualInfo.Oven.GasFlow
Info.ActualInfo.Oven.UnitFlow
$Q sends the current values from a connected KF Oven. If no Oven is connected, the
values are empty.
.HeatTime: Heating time of sample in s.
.SampleTemp: Nominal sample temperature in °C.
.LowTemp: Lowest temperature during the sample heating time in °C.
.HighTemp: Highest temperature during the sample heating time in °C.
.GasFlow: Average gas flow during sample heating time.
read only
read only
read only
read only
read only
read only
read only
read only
read only
read only
read only
read only
read only
read only
read only
read only
read only
756/831 KF Coulometer, Instructions for Use