.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of the drift
.DCor.Value: Drift value for the manual drift correction.
4.2.2.15. Mode.Parameter.Presel.IReq
Automatic inquiries after the start of the determination. From such an inquiry, the de-
termination continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 4.2.2.56) or with &M $G, see 4.2.2.1.
.ReqTitr:
4.2.2.16. Mode.Parameter.Presel.SampleUnit
Method specific sample unit, i.e. when the method is loaded, the current unit of the
sample size is overwritten by the unit from the method.
4.2.2.17. Mode.Parameter.Presel.LimSmplSize.Status
Limit control for the sample size.
4.2.2.18. Mode.Parameter.Presel.Id1Text
Text for sample identifications.
4.2.2.19. Mode.Parameter.Presel.Cell
.Cell:
.GenI:
4.2.2.20. Mode.Parameter.Presel.Oven
If an Oven is connected, its results will be incorporated into the result report of the
Coulometer.
If there is no Oven connected via RS232, this parameter has to be on "no".
4.2.2.21. Mode.Parameter.Presel.ActPuls
Output of a pulse on the I/O line "Activate", see page 132.
4.2.2.22. Mode.Def.Formulas.1.Formula
756/831 KF Coulometer, Instructions of Use
value at start.
Mode.Parameter.Presel.SReq
Mode.Parameter.Presel.ReqTitr
Current generation starts during requests (with ON).
Mode.Parameter.Presel.LimSmplSize.LoLim
Mode.Parameter.Presel.LimSmplSize.UpLim
Mode.Parameter.Presel.Id2Text
Mode.Parameter.Presel.Id3Text
Mode.Parameter.Presel.GenI
Type of generator electrode.
Current at the generator electrode in mA. "auto" means that the current is
switched in the course of determinations.
Default: 400 mA for cells without diaphragm, auto for cell with dia-
phragm.
Mode.Def.Formulas.1.TextRS
4.2 Remote control commands
id1, id1&2, all, OFF
value, unit, all, OFF
ON, OFF
g, ...up to 5 ASCII
ON, OFF
0.0...999 999
0.0...999 999
id1/C21, 10 ASCII characters
id2/C22, 10 ASCII characters
id3/C23, 10 ASCII characters
no diaph., diaphragm
100, 200, 400, auto
COM1, COM2, no
first, all, cond., OFF
H2O, CXX, RSX, +, -, *, /, (, )
8 ASCII characters
77