4.2 Remote control commands
96
Silo:
The silo memory is deleted. Same function as delete entire silo.
Assembly: All values under &Assembly.
Setup:
All values under &Setup.
All:
Values of the entire tree (except silo and method memory).
The action must be triggered with &Setup.Initalise $G.
4.2.2.99. Setup.RamInit
Initializes instrument, see page 110. All parameters are set to their default value and
error messages are cleared. The user and silo memories will be deleted.
Command only possible in the inactive state of the Coulometer.
4.2.2.100. Setup.InstrNo
Setup.InstrNo.Value
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
4.2.2.101. Diagnose.Report
Output of the report containing the adjustment parameters. The Coulometer has to be
in its inactive basic state.
4.2.2.102. Diagnose.Simulation.Keycode
0
8
16
1
9
17
2
10
18
3
11
19
4
12
20
28
5
25
21
13
27
26
29
24
4.2.2.103. Diagnose.ScreenDump
The content of the 756 Screen will be dumped to the COM which is given for manual
reports (key <CONFIG>, >peripheral units). A screen dump onto the internal printer
is not possible.
4.2.2.104. Diagnose.IntPrinter.HeatTime
Diagnose.IntPrinter.MotorSpeed
Settings for the internal printer.
serial number, 8 ASCII characters
Entering a keycode is like pressing the corre-
sponding key. The keys have the following key-
codes:
756/831 KF Coulometer, Instructions for Use
$G
$G
$G
0...29
23
15
6
22
$G
1...4.0...10
2...3.0...9