4.2 Remote control commands
4.2.1 Overview
The internal object tree can be divided into the following branches:
&
├
Mode
├
UserMeth
├
Config
├
SmplData
├
HotKey
├
Info
├
Assembly
├
Setup
└
Diagnose
756/831 KF Coulometer, Instructions of Use
Sample size is out of limits.
Exit: The error message disappears on next start or on introduction of new sample
size.
Validation interval is expired.
Exit: The error message disappears on next start or clear counter with
&Config.Monitoring.Validation.Clear $G.
Service date is reached.
Exit: The error message disappears on next start or change date in
&Config.Monitoring.Service.Date.
No Oven parameters: Oven not (correctly) connected.
Exit: The error message disappears on next start. If you don't wish oven parameters
in your report, select &Mode.Parameter.Presel.Oven "no" in your method(s).
Temperature in the KF Coulometer instrument too high (>60 °C).
Exit: The error message disappears if the Coulometer temperature is below 60°C.
Transmission error from Remote Box. Unknown characters.
Exit: Rectify error and switch Coulometer off and on again.
Time-out error from PC keyboard (Remote Box)
Exit: Rectify error and switch Coulometer off and on again.
Check Remote Box. Remote Box not (properly) connected but activated in
&Config.Periph.RemoteBox.
Exit: Rectify error and switch Coulometer off and on again.
Root
Method parameters
Administration of the internal user-memory for methods
Instrument configuration
Sample specific data
Keys with direct access
Current Data
Component data
Setting the operating mode
Diagnostics program
4.2 Remote control commands
61