result:
water crv: Mass of water in ug vs. time
rate crv:
meas crv: Potential vs. Time
comb:
mplist:
param:
calc:
C-fmla:
def:
statistics: Statistics table with the individual results.
smpl data: Current sample data.
silo:
scalc full: Full report of the silo calculations.
scalc srt: Short report of the silo calculations.
config:
user method: Contents of the method memory.
all:
ff:
Reports which are sent from the Coulometer are marked with space (ASCII 32) and '
at the beginning. Then an individual identifier for each report follows.
4.2.2.64. Info.Checksums
The checksums can be used to identify the content of a file unequivocally, e.g. files
with identical content have identical results of the checksums. An empty file has
checksum "0". The calculation of the checksums is triggered with $G.
.ActualMethod: Result of the checksum of the current method in the working mem-
4.2.2.65. Info.DetermData
Determination data in hexadecimal format.
.Write:
4.2.2.66. Info.TitrResults.RS.1.Value
.RS: Values of the calculated results.
.EP: Endpoint:
.Var: Various variables. You may overwrite the variables C40...C45, see 4.2.2.66.
756/831 KF Coulometer, Instructions of Use
Result report of the last completed determination.
Rate in ug/min vs. time
Mass of water in ug & rate in ug/min vs. time
Measuring point list of the running determination.
Parameter report of the current method. During a running determination
only "live"-parameters are accessible.
Calculation report of the current method.
Contents of the <C-fmla> key.
Contents of the <def> key.
Contents of the silo memory.
Configuration report.
All reports.
Form feed on printer.
Info.Checksums.ActualMethod
ory. Identical methods with different method names have the same re-
sults of the checksum.
Info.DetermData.Write
With "ON", the following nodes can be overwritten:
&Info.TitrResults.Var.C4X (X = 0...5) and &Mode.Name.
etc., up to .9
Info.TitrResults.EP.V
Info.TitrResults.EP.Meas
Info.TitrResults.Var.C40
etc., up to .C45
Mass coordinate in ug, e.g. "10.3"
Potential coordinate in mV e.g. "43.7".
4.2 Remote control commands
$G
read only
$G
ON, OFF
read only
read only
read only
read only/read+write
85