Clear clears the change information. For the output, there is a conversion from binary
to decimal, e.g.
Line No.
Output: 2
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also page 131):
Inputs:
0
1
2
3
4
5
6
7
4.2.2.70. Info.ActualInfo.Assembly.CyclNo
$Q sends the current values.
.CyclNo:
.I:
.Meas:
.Pot:
.IPulse:
.Bur.V:
.Bur.Clear: $G clears the volume counter.
4.2.2.71. Info.ActualInfo.Titrator.CyclNo
756/831 KF Coulometer, Instructions of Use
| 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 |
| 13| 12| 11| 10| 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
1
+ 2
3
= "10"
Start (pin 21)
Stop (pin 9)
Enter (pin 22)
pin 10
pin 23
pin 11
pin 24
pin 12
Info.ActualInfo.Assembly.I
Info.ActualInfo.Assembly.Meas
Info.ActualInfo.Assembly.Pot
Info.ActualInfo.Assembly.IPulse
Info.ActualInfo.Assembly.Bur.V
Info.ActualInfo.Assembly.Bur.Clear
Cycle number of the voltage measurement cycle, e.g. "127". From the
cycle number and the cycle time (see 4.2.2.77), a time frame can be set
up. The cycle number is set to 0 on switching on the instrument and on
every start. It is incremented as long as the instrument remains switched
on.
Total charge in mA·s, e.g. "667.48".
Measured value in mV at the indicator electrode, e.g. "104.2".
Voltage at generator electrode.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means
>28 V.
Current of actual pulse.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
Dosed volume of connected Dosino in ml, e.g. "5.234".
Info.ActualInfo.Titrator.Water
Info.ActualInfo.Titrator.Meas
Info.ActualInfo.Titrator.dWaterdt
4.2 Remote control commands
Outputs:
0
Ready (pin 5)
1
Cond. ok (pin 18)
2
Titration (pin 4)
3
EOD (pin 17)
4
not used (pin 3)
5
Error (pin 16)
6
Activate, line L6 (pin 1)
7
Pulse for recorder (pin 2)
8
Connected remote box (pin 6)
9
not used (pin 7)
10
not used (pin 8)
11
Change reagent (pin 13)
12
Smpl size out (pin 19)
13
Result out (pin 20)
read only
read only
read only
read only
read only
read only
$G
read only
read only
read only
read only
87