4.2 Remote control commands
80
Monitoring of reagent live time.
.Determ:
Number of determinations.
.DCounter: Counter of determinations already carried out.
.MaxTime: Maximum live time of reagent in days.
.TCounter: Time already elapsed since last reagent change.
.ReagCap: Reagent capacity in mg water.
.RCounter: Counter of reagent capacity.
.ClearCount: Clears all above counters.
.Drift:
Stable drift in ug/min.
4.2.2.32. Config.Monitoring.Change
Config.Monitoring.Change.Status
Config.Monitoring.Change.WaitTime
Config.Monitoring.Change.AspVol
Config.Monitoring.Change.SolventVol
Config.Monitoring.Change.Rinse
Config.Monitoring.Change.NoRinse
Changing of reagent. With a connected Dosino, the reagent is changed with
&Config.Monitoring.Change $G. The nod &Config.Monitoring.change.Status has to
be ≠ OFF.
Parameters for automatic reagent change:
.WaitTime: Waiting time in s after switching off the stirrer.
.AspVol:
Volume in ml of used reagent to be aspirated.
.SolventVol: Volume in ml of new reagent to be added.
.Rinse:
Volume in ml of rinsing reagent.
.NoRinse: Number of rinsing cycles.
4.2.2.33. Config.Monitoring.Validation.Status
Config.Monitoring.Validation.Interval
Config.Monitoring.Validation.Counter
Config.Monitoring.Validation.ClearCount
Monitoring of validation.
.Interval:
Time interval in days for validation.
.Counter:
Time counter in days since last validation.
.ClearCount: Clears the above counter.
4.2.2.34. Config.Monitoring.Service.Status
Config.Monitoring.Service.Date
Monitoring of service interval.
4.2.2.35. Config.Monitoring.DiagRep
Printing of system test report after each switching on of the Coulometer.
4.2.2.36. Config.PeriphUnit.CharSet1
Config.PeriphUnit.CharSet2
Selection of the character set and the graphics control characters for COM1 resp.
COM2 of the Coulometer.
auto, man., OFF
0...999 999
0...100...9999
0...100...9999
0...9999
ON, OFF
1...365...9999
0...9999
ON, OFF
XXXX-XX-XX
ON, OFF
Epson, Seiko, Citizen, HP, IBM
756/831 KF Coulometer, Instructions for Use
$G, $S
1...9
$G