SAFETY DEVICES

The device is equipped with safety, pneumatic and mechanical systems:
If the oxygen (O
) pressure is reduced, the nitrous oxide (N
2
A built-in mixing system regulates the N
the gas mixture.
If the flow of the gas mixture is interrupted, the patient sucks in air from outside via a safety valve.
Should the flow change direction back towards the device, a one-way valve protects the patient from
breathing in exhaled gases and accumulating carbon dioxide (
The FLUSH button, when pressed, interrupts the flow of N2O and generates a high flow of pure oxygen.
THE DEVICE IS EQUIPPED WITH ELECTRONIC SAFETY DEVICES THAT MONITOR AND SIGNAL
THE ANOMALIES IN THE SYSTEM.
IT IS ESSENTIAL THAT THE OPERATOR DURING ADMINISTERING:
IS ALWAYS PRESENT AND KEEPS AN EYE ON THE PATIENT'S VITAL
FUNCTIONS
KEEPS FLOWS UNDER CONTROL
CHECKS THE BAG FOR OPTIMAL PATIENT VENTILATION
EXCESSIVE BRIGHTNESS MAY MAKE IT DIFFICULT TO READ THE DISPLAY. CHECK
THAT THE % AND FLOW DELIVERED ARE AS DESIRED.
3.4

DEVICE MOVEMENT

To move the device
Check the correct positioning of accessories (circuits, masks, etc.)
Check that the power cable is not connected and positioned correctly.
Check that the cylinders are closed.
Unlock the 4 wheels and use only the handles on the sides of the cabinet for movement.
Do not move the cabinet
without using the handles
provided.
Once the movement is complete, secure the device by locking the wheels.
O in such a way that a minimum of 30% O
2
Do not climb on top of the
cabinet.
9 - 33
O) supply is automatically interrupted.
2
).
CO2
Do not sit or place weights on
the cabinet.
ENGLISH
is always present in
2