If several inhalation solutions are to be used one after the
other:
• Rinse the nebuliser out with drinking water between the indi-
vidual applications.
• Shake excess water out of the nebuliser.
• Fill the nebuliser with the next inhalation solution as
described.
3.3 Performing treatment
All of the safety instructions in these instructions for use
must have been read and understood before any treatment
is carried out.
Always hold the nebuliser upright during treatment.
Proceed as follows in order to carry out the treatment:
• Insert the connection tubing of
the nebuliser with a slight twist
into the air connection on the
compressor.
DANGER! Life-threaten-
ing situation caused by mix-
ing up tubes! If tubing sys-
tems for other devices are
present close by (e.g., for infusions), check carefully to en-
sure that the other end of the connection tubing connected
to the compressor is connected to the nebuliser. Otherwise,
there is a danger that different possible connections may be
confused with each other.
– 22 –