Pari COMPACT2 Mode D'emploi page 129

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 201
USING A STANDARD THERMAL DISINFECTOR FOR BABY
BOTTLES (NOT A MICROWAVE OVEN)
CAUTION
Risk of infection due to inadequate disinfection
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Make sure that the disinfector is clean and operating prop-
erly before every disinfection process.
• Allow the disinfection to continue until the disinfector
switches off automatically or the minimum disinfection
time stated in the instructions for use of the disinfector has
elapsed. Do not switch the device off prematurely.
Use a thermal disinfector with a runtime of at least 6 minutes.
Regarding the performance of the disinfection, the duration of
the disinfection procedure and the quantity of water required,
follow the instructions for use of the disinfector you are using.
Drying
After each cleaning and disinfection, place all product compon-
ents on a dry, clean and absorbent surface and let them dry
completely.
Cleaning the compressor housing
NOTE
Danger of device fault due to liquid penetration
If liquids get into the interior of the compressor, this may
cause a fault in the device.
• Never immerse the compressor in water.
• Never clean the compressor in running water.
• Never spray any liquids onto the compressor or the power
cord.
• If liquid does get into the compressor, it must not be used
under any circumstances. Before starting the compressor
again, contact the manufacturer or dealer.
– 29 –

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

152

Table des Matières