SICK PBS plus Notice D'instruction page 55

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operating Instructions
Fault
No output signal
No output signal
No/Incorrect output signal
Constant output signal during change
in pressure
Deviating zero point signal
Signal range too small
Signal range too small
Signal range falls
Signal range falls/too small
Warnings and errors
Device-internal warnings and errors are output via the digital display. The device continues to
measure if there are pending warnings. The output signal and the switching outputs continue to be
output.
The following table shows the codes and their meaning.
Warning
ATT1
ATT2
ATT3
ATT4
8024946 | SICK AG
Subject to change without notice
6. FAULTS
Possible cause
Line break
No/Incorrect auxiliary energy
Wiring error or switchover of switching logic/analog signal
Mechanical overload due to excessive pressure
Overload pressure limit exceeded
Mechanical overload due to excessive pressure
Auxiliary voltage too high/low
Moisture has entered the device
Membrane damage, e.g. caused by abrasive/aggressive
media, corrosion to membrane/process connection
Description
When the switching point was changed, the reset switching point was automatically
lowered by the smallest hysteresis of 0.25% by the device.
Zero point offset, prevailing pressure outside the limits
Password entry for menu access faulty
Unit cannot be shown on the display, previous unit is used
Measure
Check passage
Correct auxiliary energy
Observe pin assignment
Check output configuration
Observe pin assignment
Exchange device; consult with
the manufacture in the event
of repeated failure
Permissible overload pres-
sure limit upheld
Exchange device; consult with
the manufacture in the event
of repeated failure
Correct auxiliary energy
Mount cable correctly
Contact manufacturer and
exchange device
O P E R A T I N G I N S T R U C T I O N S P B S p l u s
PBS plus
5 5

Publicité

Table des Matières
loading

Table des Matières