Télécharger Imprimer la page

Agilent Technologies TV 401 Notice D'emploi page 238

Publicité

17
Technical Information
Window-Meanings
N.
126
130
200
201
202
203
204
205
206
210
300
301
302
320 to
399
400
238/266
Read/
Data
Write
Type
R/W
N
Vent valve opening delay
(expressed in 0.2 sec)
Reserved to Agilent service
R
N
Pump current in
mA dc
R
N
Pump voltage in Vdc
R
N
Pump power in W (pump
current x pump voltage duty
cycle)
R
N
Driving frequency in Hz
R
N
Pump temperature in °C
R
N
Pump status
R
N
Error code
R (see
N
Actual rotation speed in Hz
note 3)
R
N
Cycle time in minutes
(zeroed by the reset
command)
R
N
Cycle number (zeroed by the
reset command)
R
N
Pump life in hours (zeroed
by the reset command)
Reserved to Agilent service
R
A
CRC EPROM (QE)
TV 401/301 MS User Manual / 87-901-013-01 (B)
Description
0 to 65535
(corresponding to 0 to
13107 sec)
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see
the following figure
0 to 1010
999999
9999
999999
QE8XXXX
(where "XXXX" are
variable)
Admitted Values

Publicité

loading

Ce manuel est également adapté pour:

301 msTv 401/301/21 nsf