Agilent Technologies TV 401 Notice D'emploi page 237

Publicité

N.
108
109
110
111
117
120
121
122
123
124
125
TV 401/301 MS User Manual / 87-901-013-01 (B)
Read/
Data
Write
Type
R/W
N
Baud rate
W
L
Pump life/ cycle time/ cycle
number reset
R/W
L
Interlock type
(default = 1)
R/W
L
Analog output type: output
voltage signal proportional
to frequency or power
R/W
N
Low Speed frequency (Hz)
R/W
N
Rotational frequency setting
(Hz)
R/W
N
Maximum rotational
frequency in Hz (active only
in Stop condition)
R/W
L
Set vent valve on/off
(on = closed)
Reserved to Agilent service
R/W
L
Set the vent valve
operation
Technical Information
Window-Meanings
Description
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
750 to "Maximum
rotationl frequency"
(win 121)
(default = 750)
750 to "Maximum
rotationl
frequency" (win 121)
(default = 1010)
750 to 1010
(default = 1010)
On = 1
Off = 0
(default = 1)
Automatic = 0 (see
note 1.)
On command = 1 (see
note 2)
17
Admitted Values
237/266