PRODUCT PARAMETERS
Weight, Foot Plate
Operating forces of buttons
Battery (voltage output)
Charger Input (voltage supply)
Charger Output/Charging Time
Audio/Visual Low Battery Alarm
Motor Safety Devices
*Approx. Lifts per Charge (working ability)
Max. current input
Operating temperature
Air humidity
Sound pressure
Atmospheric pressure
Protection class, control unit
Protection class, hand control
Protection class, motor
Insulation class
Intermittens
Battery capacity
Manual emergency lowering
Electric emergency lowering/lifting
Expected lifetime
Lifting Speed
*NOTE: Varies depending upon load and stroke.
Roze™Stand Assist Patient Lift
24V DC
mb
100-240V AC ~ 50/60 Hz
29.5V DC Max 6 hours
Anti-Entrapment
*100-200 Cycles per charge
20% to 90% @ 30 not condensing
700 to 1060 hPa
10%, max, 2 minutes/18 minutes
The Roze lift passes the "velocity of lifting and lowering" in the
ENISO10535 (<0.15 m/s under maximum load and <0.25 m/s
8
6 kg/13 lbs
Max. 5N
max. 240VA (RCHBL)
Yes
Max. 400 mA
5° to 40° C
< 50 dB
IPX4
IPX4
IPX4
II - type B
2,9 Ah
Yes
Yes/Yes
8 years
unloaded)
Part No. 1164940