SECTION 6—TROUBLESHOOTING
SECTION 6—TROUBLESHOOTING
SYMPTOMS
Noisy or dry sound from pivots.
Electric actuator fails to lift when button is
pressed.
Unusual noise from lift arm actuator.
Lift arms will not lower in uppermost position.
Legs don't open and close properly.
Service light is flashing.
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.
Roze™Stand Assist Patient Lift
FAULTS
Needs lubrication.
Hand-control or actuator
connector loose.
Battery low.
RED emergency stop button
pressed IN.
Battery not connected properly
to control box.
The connecting terminals are
damaged.
Electric actuator in need of ser-
vice or load is too high.
Actuator is worn or damaged
or spindle is bent.
Lift arms require a minimum
weight load to lower from the
uppermost position.
Leg actuators may be worn or
damaged or spindle is bent.
12 months have passed since
the lift was manufactured or
initially assembled.
24
SOLUTION
Refer to Lubricating the Lift on page 26.
Charge batteries. Refer to Secondary Emer-
gency Release on page 15.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Reconnect the battery to the control box.
Refer to Secondary Emergency Release on
page 15.
Replace the battery pack. Refer to Second-
ary Emergency Release on page 15.
Refer to Replacing the Electric Actuator on
page 27.
Refer to Replacing the Electric Actuator on
page 27.
Pull down slightly on the lift arms.
Refer to Replacing the Leg Actuators on
page 28.
Initial assembly - Reset the service light.
Refer to Resetting the Service Light on
page 12.
After initial assembly - Service the lift. Con-
tact your local Invacare dealer.
Part No. 1164940