3
Methods
0.0.0
Zero
Test
br2
R E S U L T
R E S U L T
Bromine 0.1 - 13 mg/l
Perform zero calibration (see "Operation").
Empty the vial. Hold the drip bottle vertically and add evenly sized
drops to the vial by pressing slowly (6 drops of DPD 1 buffer
solution, 2 drops of DPD 1 reagent solution). Add the water sample
to the 10 ml mark. Screw the cap on, swirl to mix, and replace the
vial in the compartment making sure the ∇ and ∆ marks are aligned.
Press the ZERO/TEST key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l bromine.
0 - 2.3 mg/l : ± 0.12 mg/l
Tolerance:
> 2.3 - 4.5 mg/l : ± 0.25 mg/l
> 4.5 - 6.8 mg/l : ± 0.45 mg/l
> 6.8 - 9 mg/l : ± 0.68 mg/l
> 9 - 13 mg/l : ± 0.90 mg/l
20