Measured value display
OPb
calibration of the sensor
P3F
Third terminal of a
Pt100 sensor
(if present)
Digital inputs
HES
Temperature increase/
Reduction
i1F
Function of the digital
Input
i1P
Polarity of the digital
input
Alarm delay time
did
of the digital input
(0...120 min)
Miscellaneous
Adr
Serial address RS485
(1...247)
Set device to STAND-BY
OnF
Ptb
Number of the parameter-
table
rEL
Version
Display of the parameters
Pr2
in Level Pr2
regardless of the measurement range
no = Pt100 2-conductor wired
yES = Pt100 3-conductor wired
Set-value is increased/lowered by HES
during the energy saving phase.
Activation via digital input
(i1F = Es).
EAL = external alarm; OFF = unused;
bAL = serious external alarm;
Es = Energy saving mode Start/Stop;
onF = switch the device ON/OFF
externally;
C-H = reverse the control effect
CL: active with closed contact
OP: active with opened contact
For i1F = EAL or i1F = bAL.
The corresponding alarm then takes
place display or message.
Address for XJ500 recording
system. Identifies the device if it
is linked into a ModBUS-compatible
system.
no = not possible via keyboard
yes = STAND-BY can be activated by
holding down the SET button for at least
4 sec. If you want to change the set-
value, hold down the SET button for 2
sec.
only read-out value
only read-out value
display only
0911 - 13