Change Invia Liberty canister
and Invia Liberty tubing
Change Invia Liberty canister minimum once a week, when canister looks full on visual inspection or
when "Canister full" / "Filter clogged" alarm signal is activated.
Change Invia Liberty tubing minimum once a week, when tubing looks clogged on visual inspection or
when "System clogged" alarm signal is activated.
➡
1.
Provide sterile canister and sterile tubing.
2.
Press
"Standby"
[
running. Clamp the canister tubing.
3.1 Position the clamp next to the quick-connector and close the
clamp on the pump tubing and close the clamp on the dressing
tubing.
3.2 Disconnect the canister tubing from dressing tubing.
3.3 Protect the end of dressing tubing from contamination.
4.
Release and remove canister.
5.
Seal used canister with a cap located on the side of the canister.
6.
Remove canister tubing in direction of the arrow.
> 3 seconds] for the pump motor to stop
➡ ➡
27