Electromagnetic compatibility (EMC, IEC 60601-1-2:2014 4th Edition,
Table 6)
Recommended separation distance between portable and mobile RF communications
equipment and the Invia Liberty pump
The Invia Liberty pump is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Invia Liberty pump can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Invia Liberty pump as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum out-
put power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommened separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
126
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
P
d = 0.35
0.035
0.11
0.35
1.11
3.5
m
80 MHz to 800 MHz
P
d = 0.35
0.035
0.11
0.35
1.11
3.5
800 MHz to 2.5 GHz
P
d = 0.7
0.07
0.22
0.7
2.21
7.0