IV. GENERAL SAFETY INFORMATION
x
Before use, check that the handpiece is not damaged
and that no part of it is missing.
Use protective gloves, glasses and a mask.
E
When inserting or changing the rotary instrument,
make sure that the handpiece is stationary.
After insertion of a rotary instrument, check that it is
correctly held by a slight rotary movement.
Do not handle the rotary instrument during operation.
Only operate the locking catch when the motor is stopped.
The handpiece should be connected or disconnected to the
motor when the motor is stationary.
In order to cool the treatment area, only an adapted and a
correctly tempered cooling spray must be used.
Always make sure that the spray is sufficient and regular.
Do not start rotating the handpiece:
• If there is no rotary tool inserted.
• Or If the ring is not in the locked position (Close).
E
In the event of malfunction or damage, immediately
stop using the instrument and inform your approved
distributor or the manufacturer.
In the event of any questions about the equipment,
contact either your approved distributor or the
manufacturer.
No alteration or addition is to be made to the product without
anthogyr's express agreement.
E
A poorly maintained instrument may overheat and
cause unusually severe burns.
V. TECHNICAL SPECIFICATIONS
Reference
Reduction / multiplication rate
Colour code
Weight (g)
Motor connection standard
Maximum motor speed (rpm)
Maximum instrument torque (N.cm)
Tool type according to ISO 1797-1
Diameter of tools (Ø mm) according to ISO
1797-1
Maximum length recommended by
Anthogyr (mm)
Maximum diameter of the active part of the
tool recommended by Anthogyr (Ø mm)
Water spray output according to ISO 7785-2
(ml/min)
* Suggested values. If using longer and wider diameter rotary
instruments, the user is responsible for the correct choice of the
operating conditions to prevent any risk to the patient or to a third
person.
10
4330D
1 : 1
blue
39 g
ISO 3964
40000
6 N.cm
Type2
2.35
45*
10*
> 50