Dentsply Sirona
Operating Instructions CEREC AF / AF Connect / AI / AI Connect / AI for Rear
66 37 487 D3652
D3652.201.01.19.02 10.2018
Performing the test
Tip: Allow the "Sirona Windows Diagnostics" test to run for at
least 1 hour. For optimum results, allow the test to run overnight.
ü You have started the "Sirona Windows Diagnostics" test
and Windows starts in the test environment. The "Sirona
Windows Diagnostics" test starts automatically.
➢ Check the result in the "Windows Stress Test" window.
Tip: You may have to rearrange the windows to be able to see the
"Windows Stress Test" window.
Test result: Pass
No errors were found on the PC-specific hardware. Replacing the PC
component or the PC is not advisable.
Test result: Fail
An error was found on the PC-specific hardware. Replacing the PC
component or the PC may be required.
An image restore is not advisable.
1. Check the "Windows Stress Test" window for those
components that did not pass the test.
2. Carry out the appropriate steps in the chapter entitled
Troubleshooting [→ 90] .
Ending the "Sirona Windows Diagnostics" test
➢ Click the "Stop" button.
Ä The individual test windows close.
Ä The PC is switched off.
10 PC Diagnostic Tool
89