Dentsply Sirona CEREC AF Notice D'utilisation page 80

Masquer les pouces Voir aussi pour CEREC AF:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

8 Maintenance
78
Operating Instructions CEREC AF / AF Connect / AI / AI Connect / AI for Rear
Measuring the position of the exit window
1. Mount the bottom side of the calibration set to the tip of the camera.
2. Click on the "OK" button.
Ä The calibration process is continued.
Ä Once the calibration is complete, a message is displayed
indicating this.
3. Confirm the message by clicking the "OK" button.
Ä The CEREC Omnicam is calibrated.
Error message during calibration
The software indicates if an error occurs during calibration. If the
calibration process resulted in errors, restart the process.
End calibration
ü The software indicates that the calibration was completed
successfully.
➢ Click on the "OK" button.
Ä The CEREC Omnicam is calibrated.
Dentsply Sirona
66 37 487 D3652
D3652.201.01.19.02    10.2018

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Cerec af connectCerec aiCerec ai connectCerec ai for rear

Table des Matières