RACCORDEMENT ÉLECTRIQUE
4

4.3.1 Connection in the terminal compartment

Procedure
• Unscrew the housing cover.
• If present, remove the display and adjustment module by turning it to the left.
• Loosen union nut of the cable gland.
• For preparation of connection cable se référer à
• Push the cable through the cable gland into the terminal compartment.
• Insert the wire ends into the open terminal connection according to the wiring plan. Flexible
cores with wire end sleeves as well as solid cores can be inserted directly into the terminal
openings. In case of flexible cores, press the spring terminal with a small screwdriver to open
the terminal opening.
• Check the proper hold of the wires in the terminals by lightly pulling on them.
• Connect the cable shield to the internal ground terminal, connect the outer ground terminal to
the customer/plant equipotential bonding.
• Tighten the union nut of the cable gland. The sealing ring must completely enclose the cable.
• Screw the housing cover back on.

4.3.2 Connection in the housing base (external housing)

Procedure
• Unscrew the housing cover
• If present, remove the display and adjustment module by turning it to the left
• Loosen union nut of the cable gland
• For preparation of connection cable se référer à
• Push the cable through the cable gland into the terminal compartment
• Insert the wire ends into the open terminals according to the wiring plan. Solid cores as well
as flexible cores with wire end sleeves can be inserted directly into the terminal openings. In
case of flexible cores, press the spring terminal with a small screwdriver to open the terminal
opening.
• Check the proper hold of the wires in the terminals by lightly pulling on them
• Connect the cable shield to the internal ground terminal, connect the outer ground terminal to
the equipotential bonding
• Tighten the union nut of the cable gland. The sealing ring must completely enclose the cable
• Screw the housing cover back on