Advanced Error Codes - KROHNE OPTIBAR PM 5060 C Manuel De Référence

Masquer les pouces Voir aussi pour OPTIBAR PM 5060 C:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

OPTIBAR PM 5060 C

5.10.1 Advanced Error codes

Failure
Code
Text message
F013
No measurement value
available
F017
Adjustment span too small
F025
Error in the linearisation
table
F36
No operable sensor
software
F40
Error in the electronics
F041
Error in the electronics
F042
Communication error slave
F080
F105
Measured value is
determined.
F125
Impermissible electronics
temperature
F260
Error in the calibration
F261
Error in the configuration
F264
Installation / start-up error
F265
Measurement function
disturbed
F266
Impermissible operating
voltage
01/2021 - 4004819802 - MA OPTIBAR PM 5060 C R04 fr
Cause
No valid measured value
available
Adjustment not within
specification
Index markers are not
continuously rising, for
example illogical value
pairs
Failed or interrupted
software update
Hardware defect
No connection to sensor
electronics
No connection to the slave
General software error
Device is still in the
energizing phase. The
measured value has not yet
been determined.
Temperature of the
electronics in the non-
specified section
Error in the calibration
carried out in the factory
Error in EEPROM
Error during setup
Error when carrying out a
reset
Inconsistent settings (e.g.
distance, adjustment units
in application of process
pressure) for selected
applications or invalid
sensor configuration (e.g.
application of electronic
differential pressure with a
connected differential
pressure measuring cell)
Sensor no longer carries
out a measurement
Operating voltage is below
the specified range
www.krohne.com
MISE EN SERVICE
Action or description
Check the measuring cell or
the overpressure or low
pressure conditions.
Change adjustment
according to the limit values
check the linearization
table,
Delete table/Create new
Repeat software update
Check electronics version
Exchange the electronics or
send device for repair
Exchange the electronics or
send device for repair
Check connection to sensor
electronics (with remote
version)
Check the connection
between the master and
slave
Disconnect operating
voltage briefly
Wait until the energizing
phase is complete.
Check ambient temperature
Isolate electronics
Use device with higher
temperature range
Exchange the electronics
Send device for repair
Repeat setup,
Repeat reset
Changing settings or
changing connected sensor
configuration or application
Carry out a reset
Disconnect operating
voltage briefly
Check electrical connection
- if necessary, increase
operating voltage
DevSpec State in CMD 48
Bit 0 of byte 0...5
Bit 1 of byte 0...5
Bit 2 of byte 0...5
Bit 3 of byte 0...5
Bit 4 of byte 0...5
Bit 5 of byte 0...5
Bit 6 of byte 0...5
Bit 7 of byte 0...5
Bit 10 of byte 0...5
Bit 11 of byte 0...5
Bit 12 of byte 0...5
Bit 13 of byte 0...5