After connection, the sensor adjustment menu appears on the respective operating device. The
display of the display and adjustment module shows the Bluetooth symbol and connected
Sensor adjustment via the keys of the display and adjustment module itself is not possible in this
mode. If the Bluetooth connection is interrupted, e.g. due to a too large distance between the two
devices, this is displayed on the operating device. The message disappears when the connection
is restored.
Sensor parameter adjustment
Sensor parameter adjustment
Sensor parameter adjustment
Sensor parameter adjustment
The sensor adjustment menu is divided into Setup
parameters and confirm via the keyboard or the editing field. The settings are then active in the
instrument. Close the app to terminate the connection.

5.5 Quick set-up

To adapt the device quickly and easily to the application, select the menu item "Quick setup".
This parameter adjustment essentially involves the selection of the application, position
correction and adjustment of the span.
INFORMATION !
Differential pressure, density and interface are only available for electronic differential pressure
when the slave sensor is activated
Figure 5-1: Quick set-up
In this chapter, not all settings are displayed graphically but all settings are described.
Figure 5-2: Overview Quick setup
1 Measurement loop name
Measurement loop name
Measurement loop name
Measurement loop name
Assign a suitable measurement loop name
2 Electronic differential pressure
Electronic differential pressure
Electronic differential pressure
Electronic differential pressure
Activate/deactivate the slave sensor for electronic differential pressure. If you have not connected a slave sensor, con-
firm this with "deactivate".
Application
Application
Application
Application
Then select the application. The selection includes process pressure, level, differential pressure, density and interfa-
ce.
3 Adjustment units
Adjustment units
Adjustment units
Adjustment units
Determine the adjustment and temperature units of the device.
01/2021 - 4004819802 - MA OPTIBAR PM 5060 C R04 fr
Setup and Diagnostics
Setup
Setup
Diagnostics. Enter the requested
Diagnostics
Diagnostics
www.krohne.com
MISE EN SERVICE
connected.
connected
connected
5