Item
Description
Calibration
Displays the Calibration screen where you enter a blood glucose (BG) meter read-
ing for sensor calibration. For details, see Calibrating your sensor, on page 8.
The color of the ring and the size of its displayed segment indicate calibration sta-
tus and the approximate time left until your next sensor calibration is due. When
your sensor is fully calibrated, the ring is solid green. As the time for your next sen-
sor calibration approaches, segments of the ring disappear, and the color of the
ring changes, until a red blood drop appears instead of the ring, as shown below:
A full green circle indicates 12 hours remaining to the next calibration. Yellow indi-
cates three hours remaining. Red indicates one hour remaining. When the red
blood drop appears, an alert is displayed instructing you to calibrate your system.
The question mark indicates that the calibration status is unknown. Three dots indi-
cate that a calibration currently is not permitted.
Rate of change
Display your glucose trend and the rate at which the most recent sensor glucose
arrows
level has risen or fallen. For details on setting the rate of change for falling and ris-
ing sensor glucose values, see Setting rate alerts, on page 14.
Menu screen
Provides access to the menu screen, which displays the Guardian Connect Sys-
tem status information and allows you to set up your sensor, define alert settings,
view the logbook, and access CareLink Personal settings. For details, see Menu
screen, on page 6.
Continuous sen-
Displays your current and previous SG readings. Tap any point on the graph to
sor glucose
view the details of the selected sensor glucose reading or event in a box that ap-
trace
pears above. For more navigating tips, see Navigating through the sensor glucose
graph, on page 15.
System status
Display the corresponding status icon if the transmitter battery is empty, the life of
icons
your sensor has expired, the connection with the transmitter is lost, or the status of
these features is unknown. For details, see System status icons, on page 7.
Time interval
Displays preset time intervals of 3, 6, 12, and 24 hours. Tap this icon to switch be-
tween time intervals.
Current sensor
Displays your current sensor glucose reading, which the transmitter calculates and
glucose reading
sends wirelessly to the Guardian Connect app.
High and low
Display a line showing the high and low sensor glucose limits on the sensor graph.
sensor glucose
The orange line indicates your sensor glucose high limit; the red line indicates your
alert limits
sensor glucose low limit.
Sensor status
Displays the latest active status notification. If a current sensor glucose reading is
message
not available, the sensor status message will appear where the sensor glucose in-
formation would be displayed.