4. Once locked, the joints can be released in one of two ways:
a.
Depress the Power Switch to Depressurize Position (Figure 2).
b.
Press the Set Up Button (Figure 1 Item # 8)
Note: The SPIDER2 Set Up Button is disabled when the Switch Drape and/or Foot Pedal are
plugged in.
Figure 2. Power Switch
Instructions for Use
SPIDER2 Attachment to O.R. Table Rail
1. If SPIDER2 has been turned on and joints are locked, holding the Set Up Button allows
temporary repositioning of the SPIDER2. When Set Up Button is released, all joints will
lock again.
2. Open SPIDER2 Clamp by turning Clamp Actuating Knob counter clockwise so clamp
opening is larger than table rail.
3. Place Clamp over rail at an angle, placing the top jaw of the Clamp on the rail first then
lower bottom jaw into place (Figure 3).
Figure 3. L
p onto
ower Clam
table rail
4. Tighten onto table rail by turning Clamp Actuating Knob clockwise (Figure 4).
Figure 4. Tighten Kn
ob
5. Verify Clamp is rigidly attached; if not, additional tightening of the Clamp Actuating
Knob may be required.
Setting up SPIDER2 for Surgery
1. Turn SPIDER2 Power Switch to the ON position.
2. Adjust angle of Dumbbell Joint (relative to table rail) by turning Clamp Indexing Handle
counter clockwise. This separates the Clamp and allows repositioning of SPIDER2 to
vertical, horizontal, or a 45° angle (Figure 5). Turn Clamp Indexing Handle clockwise to
tighten clamp.
Figure 5. SPIDER2 Vertical, Horizontal, 45°
3. The SPIDER2 Clamp Indexing Handle (Figure 1 Item #13) position can be re-oriented by
pulling up on the Clamp Indexing Handle and turning clockwise or counter-clockwise.
Once the ideal position is found, release to lock into position.
4
126-99-01 Rev. C
4. Press Set Up Button to adjust SPIDER2 to approximate setup position and check range
of motion prior to draping.
5. Insert sterile Piggy Back into SPIDER2 Distal end.
6. Drape SPIDER2 with Switch Drape (or standard Drape if only Foot Pedal is to be used).
7. Connect Foot Pedal (Figure 6), Switch Drape (Figure 7), or both, to appropriate
Connection Receptacle.
Figure 6. Connecting Foot Pedal
8. Slide Drape under Drape Clips (Figure 7) and ensure that the Drape completely covers
the receptacles to shield them from fluids.
Figure 7. Switch Drape Connected with Drape secured under Clips
WARNING: Do not allow fluids to enter the Switch Drape Connection
Receptacle or the Foot Pedal Connection Receptacle.
9. Patient's limb is attached to SPIDER2 by various accessories. Each accessory has a
male post that connects into the female receptor on the Distal end of the SPIDER2. After
inserting accessories, lock Quick Connect to avoid accidental disconnection.
10. Position Switch Drape Release Button (if used) in convenient location for easy access
throughout procedure.
11. Refer to the desired accessory's Instructions for Use for set up and positioning
instructions.
Positioning using the SPIDER2
1. Ensure power switch is in ON position; Indicator light should be lit solid. If blinking, or
OFF, replace battery.
2. Grasp patient's limb (attached to SPIDER2) and/or SPIDER2 slender arm.
3. Press Switch Drape Release Button or step on Foot Pedal; this immediately releases all
three SPIDER2 joints.
4. Move patient's limb to desired position.
5. Take finger off Switch Drape Release Button or foot off Foot Pedal; this immediately locks
all three SPIDER2 joints.
To Remove SPIDER2 from O.R. Table Rail
1. Turn the Clamp Actuation Knob counterclockwise to release jaw from table rail.
2. Tilt SPIDER2 over the rail and lift.
3. If Clamp does not release rail after Clamp Actuating Handle has been turned
counterclockwise, tap the top of the Clamp Actuating Handle; Clamp should then
release the rail.
Note: Do not store SPIDER2 in locked position as this will drastically reduce the life
expectancy of device. Always DEPRESSURIZE before storing by disconnecting the Foot
Pedal and/or Switch Drape then pressing the Power Switch to the DEPRESSURIZE Position
for 1 second. The Power Switch will then return to the OFF/STORAGE Position (Figure 2).
SPIDER2 Limb Positioner
Instructions for Use