MIXER CONTROLLER OPERATION

Powering On

Make sure that the motor cable is plugged in and locked into the controller and the controller is
connected to the mains supply with provided power cable. Place the rear-panel POWER switch to
the ON position. If the controller is connected to mains, the display will turn on indicating power.
Internal memory permits storage of all previously set parameters including speed, timer, torque
limits, zero torque, direction of rotation, and Resume. The display will show the last mode operated.
First time the controller is powered on, display shows RPM mode with a value of 0, and motor is in
PAUSE.

Button Functions

START/PAUSE
Pressing the START/PAUSE button causes the mixer motor shaft to rotate at the set speed.
Pressing the START/PAUSE button when motor shaft is rotating stops the motor.
RPM/TIMER/TRQ
Changing display modes (RPM, TIMER, TORQUE) is accomplished by pressing the desired mode
button once. The lit LED below the digital display indicates which mode is currently in the display.
Additional presses of the mode button advance to set the desired parameter.

Set Motor Speed

The RPM display mode is active when RPM LED is illuminated. The digital display will show actual
mixer speed. If the mixer is in PAUSE the display shows 0. The RPM mode precisely sets the
output speed of the shaft, as low as 50 rpm or as high as 2500 rpm. Speed can be set both when
the mixer is in operation and in PAUSE.
When in the RPM display mode, press the RPM button. This causes the display to flash and
1.
show the set speed, indicating the set speed can be adjusted.
Use the INCREASE and DECREASE buttons to set the desired speed.
2.
New set speed will flash for five seconds, then be accepted as new speed. Pressing RPM or
3.
START/PAUSE buttons will also accept the new speed. Display stops flashing and returns to
actual speed. If mixer is in operation, rotation will change to the new set speed. The RPM LED
will flash while motor is ramping to new speed.

Set Run Time

The TIMER display mode is active when MIN LED is illuminated and the digital display shows set or
remaining minutes and seconds. If display shows 00:00 the mixer is set for continuous operation.
The digital display will show the set countdown time to a maximum of 99 minutes and 59 seconds.
Countdown begins when the START/PAUSE button is pressed to start mixer operation. Pressing
START/PAUSE while motor is rotating will stop rotation and timer, with the TIMER display showing
time remaining. Timer and mixing can resume when the START/PAUSE button is pressed. The
timer can be set both when the mixer is in operation and in PAUSE.
When in the TIMER display mode, press the TIMER button. The minutes of the display will
1.
start flashing, indicating minutes can be adjusted. To reset timer to 00:00, press and hold
TIMER button two seconds.
Use the INCREASE and DECREASE buttons to set minutes for timer as desired.
2.
Press TIMER button again. The seconds of the display will start flashing, indicating seconds
3.
can be adjusted. Pressing the TIMER button repeatedly will switch between minutes and
seconds.
Use the INCREASE and DECREASE buttons to set seconds for timer as desired.