DCI uncertainty
range
0 µA - 200 µA
200 µA - 2 mA
2 mA - 20 mA
20 mA - 200 mA
200 mA - 2 A
2 A - 20 A
ACI uncertainty
range
1 µA - 200 µA
200 µA - 2 mA
2 mA - 20 mA
20 mA – 200 mA
200 mA – 2 A
2 A - 20 A
range
1 µA - 200 µA
200 µA - 2 mA
2 mA - 20 mA
20 mA - 200 mA
200 mA - 2 A
2 A - 20 A
When option 130-50, 140-50 Current Coil is used, add uncertainty 0.3% of the set current to the value specified in above table. Output
current is multiplied by factor 50.
Auxiliary parameters
range
200 uA
maximal inductive load
400 uH
*1
THD
*1
parameter includes non-linear distortion and non-harmonic noise
Function Shape
current range:
wave form:
peak value uncertainty:
displayed values:
Minimum frequency for squarewave signals is 0.1 Hz, for all others 20 Hz.
Resistance
summary range:
Resistance uncertainty
resistance range
0 Ω - 100 Ω
100 Ω - 400 Ω
400 Ω - 2 kΩ
2kΩ - 10 kΩ
10 kΩ - 40 kΩ
40 kΩ - 200 kΩ
200kΩ -1 MΩ
1 MΩ - 4 MΩ
4 MΩ - 20MΩ
20 MΩ - 50MΩ
Manuel de l'utilisateur
% value + % range
0.05 + 0.0 + 20 nA
0.02 + 0.005
0.01 + 0.003
0.01 + 0.003
0.015 + 0.005
0.02 + 0.010
% value + % range
max voltage Vef
20 Hz - 1 kHz
0.15 + 0.0 + 20 nA
0.07 + 0.01
0.05 + 0.005
0.05 + 0.005
0.05 + 0.005
0.10 + 0.03
% value + % range
max voltage Vef
5 kHz – 10 kHz
- -
0.50 + 0.07
0.50 + 0.07
0.50 + 0.07
- -
- -
2 mA
20 mA
400 uH
400 uH
0,2%
0,2%
100 uA to 2 A
square, positive, negative, symmetrical, ramp A, ramp B, triangle
truncated sin with THD 13,45 %
0.3 % + 500 nA
peak, effective
Ω
0
to 50 M
uncertainty of value [%]
0.03 + 10 mΩ
0.015
0.015
0.015
0.015
0.015
0.05
0.1
0.2
0.5
Calibrateur multifonctions CX 1651
max. voltage V
3
3
3
3
3
1.5
% value + % range
1 kHz – 5 kHz
3
0.30 + 0.10 + 20 nA
3
0.20 + 0.05
3
0.20 + 0.05
3
0.20 + 0.05
3
1.5
- -
2
2
2
- -
- -
200 mA
400 uH
200 uH
0,2%
0,2%
Ω
current range
1 mA - 40 mA
400 µA - 20 mA
100 µA - 4 mA
20 µA - 1 mA
4 µA - 200 µA
1 µA - 40 µA
0.2 µA - 10 µA
40 nA – 2 µA
10 nA – 500 nA
4 nA – 150 nA
max voltage Vef
3
3
3
3
- -
- -
- -
- -
2 A
10 A
100 uH
0,2%
0,3%
175