Calibrateur multifonctions CX 1651
42. Switch activity .... xx PASS/FAIL
This parameter controls the function of the built-in relay. If PASS is set, the relay is active (changes its status) if
the test results in PASS status. If FAIL is set, the relay is active (changes its status) if the test results in FAIL
status.
43. Cal.code .... 00000
Entry of calibration code. Calibration code is a five-digit number, which must be entered to access the calibration
mode. If the calibration code is set to "00000", this information is displayed in the Setup menu. Calibration code
can be changed. New calibration code can be directly entered using numeric keyboard and confirmed by pressing
ENTER. If non-zero calibration code is set, correct calibration code must be entered to access the calibration
mode. Non-zero calibration code is not displayed further on the display.
The purpose of the calibration code is to prevent unauthorized users from changing the calibration of the
instrument.
Note
It is advisable to write down actual calibration code if changed. If you forget the calibration code, you have to
send the calibrator to the manufacturer.
44. Cal.date .... xx.yyyy
Displays the date of last calibration of the calibrator (month/year).
45. Serial No .... xxxxxx
46. Time .... xx:yy
47. Date .... xx.yy.zzzz
48. Time on display .... xx
ON/OFF
144