The display shows the following data:
*
main data of temperature in
*
sensor type
o
*
resistance at 0
C labeled R0 (resistance temperature sensors only)
*
cold junction temperature of thermocouple sensors labeled RJ
set value of relate deviation in %, labeled ∆T = xxxx.x °C (K)
*
the information section shows:
*
temperature scale type
*
uncertainty of simulated temperature value of selected temperature sensor type
•
Set the main value of temperature using numeric keyboard, cursor buttons or potentiometer. Output terminals
are disconnected, the information section of the display shows the symbol
terminals are disconnected.
•
Connect the object to be calibrated to Hi - Lo terminals.
•
Press OUTPUT button.
•
Red LED is lit above the OUTPUT terminal to indicate that the output signal is connected to output
terminals. The display shows the symbol of connected output terminals.
Note
•
Load of output terminals is limited similarly to corresponding voltage or current ranges.
•
Output signals provided at Hi - Lo terminals and AUXILIARY connector are short-circuit proof.
Switching between temperature sensor types
•
Keep pressing "TC type" or "RTD type" display button to select desired sensor type.
•
If resistance temperature sensors are selected, each press of the button selects Pt1.385, Pt1.392 or Ni
resistance thermometer. The display shows current setting as Pt385 / Pt392 / Ni.
•
If thermocouples are selected, each press of the button selects K, N, R, S, B, J, T, E types. The display shows
current setting as TC TYPE x, where x is the type of the thermocouple.
Manuel de l'utilisateur
o
C or K
thermocouples:
resistance temperature sensors:
Calibrateur multifonctions CX 1651
K, N, R, S, B, J, T, E
Pt 1.385, Pt 1.392, Ni