When the multimeter is being set-up, measurement cannot be initiated. Measurement can only be initiated after
the menu is left using the INPUT button.

Function selection

Press METER direct control button and select FUNCTION option from the menu using cursor buttons.
Press UP, DOWN display buttons or use the potentiometer to select one of the following functions:
*
basic DC voltage range VDC 10 V
*
DC current mA DC 20 mA
*
small DC voltage range mV DC up to 2 V
*
four-wire resistance measuring range up to 2 kΩ
*
frequency up to 15 kHz
*
temperature measurement using thermocouples T TC
*
temperature measurement using resistance temperature sensors T RTD
*
measurement of strain gauges (pressure, force) SGS
By selection of the measurement function, some related options are automatically changed as well
(measuring range, unit of measurement).
Press EXIT display button the close the function selection menu and to return to the basic display.

Setting the measurement range

Press METER direct control button and select RANGE option from the menu using cursor buttons.
Press UP, DOWN display buttons or use the potentiometer to select the measurement range for selected
function. Ranges of individual functions can be changed:
*
small DC voltages mV DC ranges:
*
temperature measurement using thermocouples T TC, types:
*
temperature measurement using RTD sensors, types:
*
setting of sensitivity of strain gauge sensors (pressure, force) SGS
Manuel de l'utilisateur
Calibrateur multifonctions CX 1651
20, 200, 2000 mV
K, N, R, S, B, J, T, E
Pt 1.385, Pt 1.392
131