Testing of regulation and measurement sets and evaluation units

The calibrator can be used for calibration and testing of various instruments and regulators which must be fed by
precise signal and whose response (electrical signal) is to be measured.
Two methods of connection can be used, each of them having different function and generation and measurement
ranges.
Simultaneous mode can only be used if Opt 40, 60, or 140-41 cable adapter is connected to AUXILIARY
terminal. When no adapter is connected, simultaneous mode cannot be activated.

Use of Opt. 140-41 cable adapter

This adapter allows full utilization of the features of the built-in multimeter. It can be used during the calibration
of objects which require the simulation of temperature or other non-electrical sensor signals, or the generation of
small DC voltages at the input side, and the measurement of standard outputs (current loop, standard voltage) at
the output side.
When using 140-41 cable adapter, OUTPUT 140-41 option must be set in the setup menu. This function selects
the calibrator's output terminals. If OUTPUT 140-41 option is set to PANEL, calibrator's output is connected to
front panel terminals (not to the cable adapter). If this option is set to AUX, output signals are connected to the
cable adapter (not to the front panel terminals). "Multimeter/Simultaneous function" chapter defines the limits of
functions and ranges.
If 140-41 cable adapter is used for simultaneous calibration and OUTPUT 140-41 option is set to AUX, the
following generation/measurement ranges can be set:
Generator
DCV
DCI
Resistance
RTD sensor simulation
TC sensor simulation
Meter
DCV
mVDC
DCI
Resistance
Frequency
170
0 – 20 V
0 – 20 mA
0 – 50 MΩ
Pt 1.385, Pt 1.392, Ni
K, N, R, S, B, J, T, E
0 – 12 V
0 – 2 V
0 – 25 mA
0 – 2 kΩ
1 – 15kHz