Select the value to be changed by cursor buttons or the knob of the potentiometer.
Enter the numeric value using the units of measurement displayed in the header of the table.
Confirm the entry by pressing ENTER. New entry is copied to the respective position.
This way, all numeric entries in the table can be changed.
Press EXIT display button to return to the previous level when the programming is complete.
Note:
When the test program is executed, all steps are performed.
Within one test program, only one type of input signal and output signal can be used. It is impossible to switch to
other function of the calibrator in the middle of a program.

Setting the relays

SWITCH POLARITY and SWITCH ACTIVITY items of SETUP MENU are used to set-up the relay. The
following table lists the states of the relay depending on the values of SWITCH POLARITY, SWITCH
ACTIVITY and the test result.

SETUP MENU

Switch polarity
OFF
Switch activity
PASS
Switch polarity
ON
Switch activity
PASS
Switch polarity
OFF
Switch activity
FAIL
Switch polarity
ON
Switch activity
FAIL
Setup menu
Manuel de l'utilisateur
RELAY BEFORE TEST
disconnected
connected
disconnected
connected
Calibrateur multifonctions CX 1651
TEST RESULT
RELYA AFTER TEST
FAIL
PASS
FAIL
PASS
FAIL
PASS
FAIL
PASS
disconnected
connected
connected
disconnected
connected
disconnected
disconnected
connected