Télécharger Imprimer la page

Use Of Option 40/60 Cable Adapter; Use Of Option 70; Examples Of Tests - Metrix CX 1651 Manuel De L'utilisateur

Calibrateur multifonctions

Publicité

Les langues disponibles

Les langues disponibles

If 140-41 cable adapter is used for simultaneous calibration and OUTPUT 140-41 option is set to PANEL, the
following generation/measurement ranges can be set:
Generator
DC/AC V
DC/ AC I
Resistance
RTD sensor simulation
TC sensor simulation
Fequency
Power/Energy
Meter
DCV
MVDC
DCI
Resistance
Frequency
When 140-41 cable adapter is connected, only harmonic signals are generated (SHAPE function is inactive).

Use of Option 40/60 cable adapter

Opt. 40 cable supports simultaneous mode - measurement of voltage up to 12 V, current up to 25 mA and
frequency up to 15 kHz. Opt. 60 cable is designed for four-wire measurement of resistance or temperature using
resistance temperature sensors. The generation ranges are the same as when 140-41 cable adapter is used.

Use of Option 70

Adapter is aimed for four-wire generation of resistance values and for four-wire resistance temperature sensor
simulation. Re-calibration of resistance function of the calibrator can be performed only with this adapter.
Resistance ranges are the same as two-wire resistance ranges with output on Hi – Lo terminals on front panel.
When adapter Option 60 is used, resistance output is only on output terminals on this adapter. Front panel Hi –
Lo terminals are switched off.
When adapter Option 60 is used, set resistance value is connected to the adapter output terminals permanently.
Status of button ON/OFF doesn't influence resistance connecting or disconnecting.

Examples of tests

Examples of applications
Generation
Measurement
Pt 100
10 V / 20 mA / f
TC
10 V / 20 mA / f
Frequency
10 V / 20 mA
Resistance
10 V / 20 mA
Manuel de l'utilisateur
0 – 1000 V
0 – 20 A
0 – 50 MΩ
Pt 1.385, Pt 1.392, Ni
K, N, R, S, B, J, T, E
0.1 – 20 MHz
0.2 – 240 V / 0.2 – 20 A
0 – 12 V
0 – 2 V
0 – 25 mA
0 – 2 kΩ
1 – 15 kHz
Application
Calibration of temperature evaluation units, setting of
temperature regulators
Calibration of temperature evaluation units, setting of
temperature regulators
Setting/calibration of energy meters
Measurement of resistance bridges
Calibrateur multifonctions CX 1651