Calibrateur multifonctions CX 1651

Multimeter

The calibrator includes a built-in multimeter which can measure many electrical and non-electrical values.
Besides DC voltage and current, it can measure frequency, temperature and when external strain gauge sensors is
connected, even other non-electrical values can be measured. The multimeter can only be connected through
AUXILIARY connector. „Description of controls" chapter lists the pins of the connector.
The manufacturer recommends 140-41 cable adapter or Option 40 cable terminal to connect the multimeter. 140-
41 cable adapter includes a fuse which protects the multimeter from current overload.
The manufacturer does not recommend connection of multimeter inputs to
Such connection can result in presence of high voltage at multimeter inputs,

Basic menu

Multimeter setting menu can be opened by pressing the METER direct control button. After the button is
pressed, SETUP METER menu opens, which allows to set-up the multimeter. Press EXIT display button to
return to previous menu.
Press ∧ or ∨ cursor button to browse the menu options. Active option is always inverted.
The parameters of the active option can be changed if the respective function is allowed to be changed. Use
UP, DOWN, CLR, NEXT and display buttons to change the parameters of the active option.
Multimeter parameters can also be set using the potentiometer. The potentiometer can be used either to
browse the menu options or to change the value of the active option. Press the knob to switch between these
modes.
Keep pressing EXIT display button to return to normal display when you are finished setting up the
parameters.
130
calibrator outputs.
which can damage the multimeter.